This item is sold by
Please sign in so that we can notify you about a reply
Description
Main performance and Technical Index
- Ventilation mode: VCV, PCV, PSV, SIMV(VCV)+PSV, SIMV(PCV)+PSV, MMV, CPAP, Backup apnea.
Air circuit Performance:
- Maximum safety pressure of air circuit: no more than 6kpa.
- Inspiratory oxygen concentration: 21-100%.
Ventilation Performance:
- Respiratory rate (f): 4-99 times / min.
- Inspiratory time (Ti): 0.2-6 s
- Breath holding time (Tp): 0-2s
- Triggering pressure (Pt): inspiratory triggering pressure is - 2kpa-0kpa
- Tidal volume (Vt): 0-2000ml
- Maximum minute ventilation volume: not less than 18 L / min.
- PSV:0~5 kPa
- Peep Pressure: 0-3kpa
- Inspiratory flow rate: 0-80 L / min.
- SIMV frequency: 6-60 times / min.
- Inspiratory pressure: 0-6kpa
Alarm Performance:
- Pressure High Limit (Ph) alarm: 2 ~ 6kpa, audible and visual alarm.
- Pressure Low Limit (Pl) alarm: 0~2KPa, audible and visual alarm.
- Minute ventilation High Limit (Mvh) alarm: audible and visual alarm when ventilation exceeds the set value.
- Minute ventilator Low limit (Mvl) alarm: audible and visual alarm when ventilation is lower than the set value.
- Power off alarm: Power off alarm immediately, and the alarm duration shall not be less than 120s.
- Mute: no more than 2 minutes.
Monitoring Performance:
- 12.1 inch color LCD
- Tidal volume (VT): 0 ~ 2000ml
- Measured total frequency (Ftot), Spont. frequency (Fspont)
- Minute ventilation (Mv).
- I: E ratio.
- Airway pressure (PAW).
- Peak pressure (Ppeak)
- Peep pressure: 0.1 ~ 2KPa
- Pressure-Time and Flow-Time waveform
Features
Brand:
RENDER
Country:
Pakistan
Model No.:
SD-H3000A
Place of Origin:
USA
Reviews